The results viewer supports the navigation between fields and viewing options.
With the navigation panel on the left, past field can be selected and reviewed in the results panel in the middle. The viewing options panel on the right allows selection of all available printouts.
With the user-defined view, eight different representations can be freely chosen. For glaucoma practices, the use of the Cluster and Polar Analysis for structure-function comparison is highly recommended.
The 7-in-1 view is the standard view of the Octopus perimeter including Grayscale, Comparisons and Probabilities representations. The Defect Curve supports identification of diffuse loss.
The both eyes view shows a left and right visual field next to each other for identification of left right symmetries. It is the standard view to assess chiasmal and post-chiasmal defects.
The single eye view displays one visual field at maximum magnification. This supports reviewing full field visual field tests like the binocular Esterman test in detail.
All Octopus perimeters can be switched to HFA mode anytime. In that mode, various representations are calculated using the formulae and representations of the HFA perimeter. The Single Field Analysis thus includes the HFA Grayscale, Standard Deviation and Pattern Deviation displays as well as MD and PSD. The HFA-mode is available for all viewing options.